Author(s): Neal, R. E. II.; Garcia, P. A.; Kavnoudias, H.; Rosenfeldt, F.; Mclean, C. A.; Earl, V.; Bergman, J.; Davalos, R. V.; Thomson, K. R.

In Vivo Irreversible Electroporation Kidney Ablation: Experimentally Correlated Numerical Models

Journal: IEEE Transactions on Biomedical Engineering, 62/2 (2015), pp. 561-569
DOI: 10.1109/TBME.2014.2360374
Request reprint  |  Tell your friend  | 

Abstract: Irreversible electroporation (IRE) ablation uses brief electric pulses to kill a volume of tissue without damaging the structures contraindicated for surgical resection or thermal ablation, including blood vessels and ureters. IRE offers a targeted nephron-sparing approach for treating kidney tumors, but the relevant organ-specific electrical properties and cellular susceptibility to IRE electric pulses remain to be characterized.Here, a pulse protocol of 100 electric pulses, each 100μs long, is delivered at 1 pulse/s to canine kidneys at three different voltage-to-distance ratios while measuring intrapulse current, completed 6 h before humane euthanasia. Numerical models were correlated with lesions and electrical measurements to determine electrical conductivity behavior and lethal electric field threshold. Three methods for modeling tissue response to the pulses were investigated (static, linear dynamic, and asymmetrical sigmoid dynamic), where the asymmetrical sigmoid dynamic conductivity function most accurately and precisely matched lesion dimensions, with a lethal electric field threshold of 575 ± 67 V/cm for the protocols used. The linear dynamic model also attains accurate predictions with a simpler function. These findings can aid renal IRE treatment planning under varying electrode geometries and pulse strengths. Histology showed a wholly necrotic core lesion at the highest electric fields, surrounded by a transitional perimeter of differential tissue viability dependent on renal structure.